Sampling for Fungal Aerosols: Principles of Sample Collection
By Dr. Harriet Burge, EMLab P&K Chief Aerobiologist and Director of Scientific Advisory Board
Introduction
Collection of interpretable data depends on many factors, not the least of which is the type of
sampling equipment used. Choice of an air sampler depends on the hypothesis being tested,
expected concentrations, and the nature of the analytical method to be used. Also to be considered
is the expected particle size range of the aerosol and the acceptable amount of error. These and
other factors related to the available sampling methodologies are the subject of this article.
Particle Collection Modalities
Inertial Impaction
Inertia is the tendency of an object to resist any change in its motion. Inertia is directly
proportional to mass. Thus, larger particles require greater force to be diverted from their
original path than do smaller particles. Inertial impaction methods for spore collection generally
increase particle inertia by accelerating an airstream closely subtended by a perpendicular
collection surface.
Fungal aerosols consist of a wide range of particle types and sizes so a consideration of the
size range of particles to be collected is essential. The size of importance is the aerodynamic
size, which is the diameter of a sphere that would have the same mass as the spore in question.
Mass is a function of density and the mass of a spore must be determined experimentally. Fungal
spores are often considered to have unit density (1 gram/cc – the density of water) because
water is a major component. The actual density of a spore depends on the amount of water present,
as well as the density of other spore components. Regardless of these considerations, unit
density is still used to estimate aerodynamic diameter.
Another problem is that many (or perhaps most) fungal spores are not spherical, and the effect
of these nonspherical shapes is difficult to predict. For spores, the smallest diameter is
considered the aerodynamic diameter. Also, many spores are not smooth, increasing drag caused
by air molecules. The effects of surface ornamentation on aerodynamic diameter would have to
be experimentally determined.
All of these factors contribute to the inevitable errors involved in aerosol sampling,
especially when the aerosol is as complex as most fungal aerosols.
In absolutely still air, gravity is the dominant force that removes particles from the air.
Particle settling depends on aerodynamic diameter, so that aerodynamically large particles fall
from the air faster than smaller ones. Over a long period of time in a sealed environment, all
spores will have fallen from the air. However, until all spores have fallen, a gravity sample
in still air underestimates the concentration of small spores. The slightest air movement will
overcome the effect of gravity on fungal spores, which will tend to follow the air stream. The
smaller the spore, the larger this effect, and small spore concentrations are seriously
underestimated in gravity samples.
One of the parameters of a sampler that indicates its relative usefulness for bioaerosol
sampling is the d50. The d50 is the aerodynamic diameter of particles
that are collected with 50% efficiency. D50 for a sampler is calculated using the
following equation:

Stk = Stokes' number describes the behavior of a particle suspended in a fluid flow
W = width of particle
ŋ = air viscosity
p = particle density
V = average air velocity in nozzle
C = Cunningham slip correction factor
A good discussion of d50 calculations is presented in Trunov et al. (2001). Note that
the published d50 for a sampler is only obtained when the sampler is operated at the
required flow rate. In the best samplers, particles with diameters above the d50 are
collected with nearly 100% efficiency. Likewise, efficiency for particles below the d50
is very low. Thus, the shape of the efficiency curve is important. The "best" samplers
have steep efficiency curves. Shallow efficiency curves mean that the 100% efficiency diameter
may be much larger than the d50, resulting in overall inefficiency for particles
above the d50.
Inertial impaction is the most commonly used method for capturing fungal aerosols. The
representativeness of the aerosol captured using inertial impaction depends on the use of a
sampler with a d50 below the aerodynamic diameter of the smallest spore expected or
of interest. Obviously, the sampler must be operated at the required flow rate with the
appropriate collection medium in place.
Rotating impactors
The Rotorod sampler uses direct impaction to capture airborne particles. Narrow rods (< 2mm
in diameter) with greased flat surfaces are whirled rapidly through the air capturing particles
that get in the way. The efficiency of collection depends on the width of the sampling surface,
with narrower surfaces collecting smaller spores more efficiently. The d50 for the
Rotorod that is commercially available today is about 20µm. It is therefore only efficient
for the collection of very large fungal spores, which account for only a small proportion of
most fungal aerosols both indoors and out. It is very useful for pollen sampling (Solomon et al. 1980).
Suction Devices
Capturing particles from a moving air stream involves two steps: 1) the particles have to be
pulled into the sampling device; 2) the particles have to be removed from the moving air stream.
Pulling Particles into the Impactor (Isodirectional and Isokinetic Sampling)
Capturing complex fungal aerosol particles in a way that accurately represents the original
population is a significant challenge. If you place a sampling orifice so that the opening is
facing into the airstream (i.e., the opening is perpendicular to the air stream and the air
flows directly into the opening) (isodirectional) and no suction is applied, the ambient
airstream will tend to flow both into and around the orifice. The particles will follow the
path of least resistance. The largest particles will tend to continue on their original path
and enter the orifice. Smaller particles will tend to flow around the orifice and be lost. If
an amount of suction is applied that matches the velocity of the air stream, both large and
small particles should be collected with equal efficiency, since no change in inertia is
necessary. This is called "isokinetic" sampling. As you increase suction velocity
over that of the ambient air, streamlines will tend to bend into the orifice and oversampling
of the smallest spores will tend to occur. Very high velocity sampling will tend to pull all
particles into the sampler. This is usually not a problem outdoors, but such samplers essentially
clean the indoor air, resulting in possibly serious underestimates of the aerosol concentration
per cubic meter of air that would be present if the sampler were not operating. In other words,
it is important that the sampler itself not change the environment. The degree of underestimation
will depend on the volume of air sampled after the particles are gone. This is because as you
continue sampling, the denominator of the equation for calculating concentration per cubic meter
of air gradually becomes larger.
If the sampling orifice is not directed into the airstream, then some particles will have to
make a turn in order to enter the sampler. If the opening is parallel to the airstream, all
particles will have to turn. With no suction, no spores will be captured. With suction, smaller
particles will be able to make the turn and larger particles will not.
Increasing particle inertia
Airflow restriction
The principles of capturing particles by acceleration through small orifices are well
illustrated by the Andersen 6-stage impactor. This sampler uses sieve plates with holes of
descending size subtended either by greased metal plates or by agar plates to capture particles
according to their aerodynamic size, which is more or less proportional to their mass and thus
inertia. As the air stream accelerates through the largest holes, all particles gain inertia,
and those particles that have gained sufficient inertia will impact on the collection surface.
So, the top plate in the series captures the largest particles. The air is then accelerated
through smaller holes and a set of particles with less mass is captured. As the hole size
decreases, acceleration through the holes increases and smaller and smaller particles are captured.
Other kinds of samplers that use airflow restriction to accelerate spores are the spore traps
and the impingers. The Air-O-Cell, the Allergenco and the Burkard spore traps are examples of
commonly used spore traps. These samplers use a single rectangular orifice to accelerate
particles that have entered the sampler. The amount of acceleration achieved is inversely
proportional to the width of the rectangle or slit. Thus, as the slit narrows, the acceleration
increases, as does the efficiency of collection for small particles. Common slit widths range
from 1-2mm. Other spore traps use circular orifices (e.g., the Cyclex) and particle collection
efficiency is proportional to the diameter of the orifice. All of these spore traps have
collection surfaces that subtend accelerating orifices. An important engineering consideration
is the distance of this collection surface from the orifice. It must be very close if small
particle capture is to occur. The shape of the inside of the orifice is also important as it
controls the shape of the airstream. Grinshpun et al. (2007) compared the Allergenco-D and the
Air-O-Cell, which essentially differ only in the slit to surface distance and the shape of the
accelerating orifice. The Allergenco-D is more efficient for the collection of small particles
than the Air-O-Cell and the efficiency curve for the Allergenco is steeper. The d50s
for the two samplers are about 2.5µm for the Air-O-Cell and about 1.7µm for the Allergenco.
Impingers use the same principles as the spore traps, but the impaction surface is under water
(or some other liquid). The AGI (All Glass Impinger) is an example of this type of sampler. In
the AGI, air is drawn through a critical orifice that ensures sonic velocity. The opening of the
orifice is placed close to the bottom of the sampler, immersed in liquid. The particles impact
on the sampler surface beneath the liquid and are then captured in the liquid. The AGI is not
commonly used in routine fungal aerosol investigations, in part because of problems with foaming,
and the fact that many fungal spores are hydrophobic and are lost in the bubbles (Lin & Li 2010).
The SKC biosampler is an impinger in which the air stream is directed tangentially against a
wetted surface. Particles that impact are then washed to the bottom of the sampler.
Centrifugation
The RCS centrifugal sampler draws air into the device using a fan with blades slanted such that
the air and entrained particles are accelerated toward the collection surface (agar) using
centrifugal force.
Cyclone samplers also use centrifugal force to capture particles, but the particles are
generally captured in a liquid that collects in the bottom of the sampler.
Trapping the Particles
Once inside the sampler, the particles must be captured onto or into some kind of medium so
that analysis is possible. Ideally, once a spore has achieved sufficient inertia to stop at
the sampling surface, it will stay there until the sample is analyzed. However, unless the
surface of the spore itself is sticky (and some are!), then contact with a hard surface will
cause the spore to deform slightly, and the resulting elastic response will cause the spore to
bounce away from the surface. The degree of bounce depends both on the nature of the collecting
surface and the nature of the particle itself.
Agar is used as an impaction surface in culture plate impactors and the RCS centrifugal samplers.
The assumption is that because the agar is soft, the particles with sufficient inertia for
impaction will embed in the agar and not bounce. This phenomenon actually depends on the
hardness of the agar surface and, as mentioned above, how many other particles have impacted
at the same site. As the agar surface dries, bounce becomes an increasing concern.
On a hard surface some kind of adhesive is usually necessary to reduce bounce. The type of
adhesive is an important consideration. Generally, silicone sealants have proven the best at
preventing particle bounce but must be used soon after application since the sealant
"cures" and becomes less sticky (Clauß et al. 2010). The widely used Air-O-Cell and
Allergenco samplers use proprietary sticky surfaces.
Liquid can also be used to capture particles that bounce. The liquid impingers are designed so
that particles bounce from the hard glass surfaces and are captured in the liquid.
The shape of the orifice also affects particle bounce. For example, the Allergenco orifice has
a straight section (a cylindrical tube with straight sides) that reduces turbulence and leads
to a relatively compact deposit, while the Air-O-Cell lacks this straight section, resulting
in spread of the spore deposit beyond the width of the slit (Grinshpun et al. 2007).
It should be noted that overloading impactor surfaces leads to increasing bounce as the surface
becomes coated with non-sticky materials. In other words, spores may be bouncing off of other
spores and particles. Concentration on impactor surfaces may also lead to analytical difficulties
such as masking during microscopic identifications and loss of culturability with the culture
plate impactors (Mainelis & Tabayoyong 2010).
Filtration
Fungal spores are readily captured on filters. Capture is caused almost entirely by interception
except for filters with very large pore sizes. Filters with pore sizes in the range of 1.5-2µm
are very effective at capturing more than 99% of airborne fungal spores. Using the largest possible
pore size will reduce resistance and allow for the use of small personal pumps at reasonably high
flow rates. If filters are to be used for microscopic analysis, they need to be made of a material
that can be cleared, such as cellulose acetate. Epifluorescence microscopy requires filters that
have a relatively smooth surface. Black filters are often used to enhance the fluorescent image.
Teflon or other smooth surface filters should be used if spores are to be washed from filters
for analysis (Deacon et al. 2009).
Filtration is an attractive alternative for bioaerosol sampling, especially when long sampling
periods are necessary, and when aerosols are expected to be extremely concentrated, such as in
many farming environments. However, it should be noted that filtration seriously reduces
culturability of some microorganisms, especially during long sampling times. In addition,
increasing concentration of spores on the filter may clog the pores and increase backpressure
on the pump, reducing flow rate. Also spores generally must be washed from filters for analysis.
Electrostatic charges may cause spores to stick to the filter holder surfaces so that these
surfaces must be washed as well. All of these factors introduce important sources of error.
Electrostatic Precipitation
Electrostatic samplers collect particles by applying a charge to collection plates and either
taking advantage of the natural charge on airborne particles or by oppositely charging the
particles, which are then collected on the plates. This type of sampling is rarely used for
fungal aerosols although there is evidence that they would be quite efficient. Yao and Mainelis
(2006) compared collection of fungal spores indoors and out for an electrostatic sampler and
the BioSampler, which is a form of the N6 Andersen sampler. The electrostatic sampler, which
took advantage of natural charges on the spores, collected many more culturable spores than
the BioSampler. This points out the fact that, although Andersen sieve plate impactors are
highly efficient at spore collection, culturability may be compromised by the force of
impaction, or by competition of adjacent spores impacting at the same site.
Virtual Impaction
Virtual impactors don't actually capture particles by impaction, but they do take advantage of
particle inertia. Particles are drawn into the device at a rate that allows the largest particles
to enter a reservoir, while the smaller ones pass by into a second chamber. Here, the airflow
is faster and the particles that bypass the reservoir are smaller, while the remaining larger
particles are captured. Successive reservoirs capture successively smaller particles. A good
discussion of virtual impactors can be found at
www.tsi.com/uploadedFiles/Product_Information/Literature/Application_Notes/ITI-051.pdf.
Comments and Conclusions
Sampling for bioaerosols is a significant challenge that requires consideration of many different
variables. Understanding the principles of particle sample collection and the variables affecting
the efficiency of sample collection is essential if realistic sampling protocols are to be designed.
In addition, while using the same sampling device for all studies is attractive for a variety of
reasons, keeping an open mind about new methods that might be more efficient, reliable, and cost
effective is an important part of professional practice. On the other hand, within a study, the
same sampler should be used throughout, and if comparisons to other studies are planned, then it
is essential to use the same sampler under the same conditions.
References:
1. Clauß M, Springorum AC, Hartung J. 2010. Effective collection of airborne micro-organisms by
direct impaction on silicone sealants-comparison of different adherent surfaces. Aerosol Science and Technology 44(11): 993-1004.
2. Grinshpun SA, Willeke K, Ulevicius V, Juozaitis A, Terzieva S, Donnelly J, Stelma GN, Brenner
KP. 1997. Effect of impaction, bounce and reaerosolization on the collection efficiency of
impingers. Aerosol Science and Technology 26(4):326-342.
3. Lin X, Reponen TA, Willeke K, Grinshpun SA, Foarde KK, Ensor DS. 1999. Long-term sampling of
airborne bacteria and fungi into a non-evaporating liquid. Atmospheric Environment 33: 4291-4298.
4. Mainelis G, Tabayoyong M. 2010. The effect of sampling time on the overall performance of
portable microbial impactors. Aerosol Science and Technology 44(1):75-82.
5. Solomon WR, Burge HA, Boise JR. 1980. Performance of adhesives for rotating arm impactors.
J. Allergy Clin Immunol 65(6):467-470.
6. Trunov M, Trakumas S, Willeke K, Grinshpun SA, Reponen T. 2001. Collection of bioaerosol
particles by impaction: effect of fungal spore agglomeration and bounce. Aerosol Science and Technology 35: 617-624.
7. TSI Inc.: How A Virtual Impactor Works
8. Yao M, Mainelis G. 2006. Utilization of natural electrical charges on airborne microorganisms
for their collection by electrostatic means. Aerosol Science 37: 513-527.